|
Gas Chromatography
Description
Chromatography
is the science of separation which uses a diverse group of methods to separate
closely related components of complex mixtures. During gas chromatographic separation,
the sample is transported via an inert gas called the mobile phase. The mobile
phase carries the sample through a coiled tubular column where analytes interact
with a material called the stationary phase. For separation to occur, the stationary
phase must have an affinity for the analytes in the sample mixture. The mobile
phase, in contrast with the stationary phase, is inert and does not interact
chemically with the analytes. The only function of the mobile phase is to sweep
the analyte mixture through the length of the column. Gas chromatography can
be divided into two categories: (1) gas-solid and (2) gas-liquid chromatography.
Gas-liquid GC, developed in 1941, is the primary GC technique used for environmental
applications. Gas-solid GC is not widely used for environmental applications.
The stationary
phase is chosen so that the components of the sample distribute themselves between
the mobile and stationary phase to varying degrees. Those components that are
strongly retained by the stationary phase move slowly relative to the flow of
the mobile phase. In contrast, components that have a lower affinity for the
stationary phase travel through the column at a faster rate. As consequence
of the differences in mobility, sample components separate into discrete bands
that can be analyzed qualitatively and quantitatively.
Gas chromatography
is the most widely used chromatographic technique for environmental analyses.
Typical Uses
Gas chromatography (GC) analysis is a widely used technique for field-based
analysis. Analysis of organic compounds is possible for a variety of matrices
such as water, soil, soil gas, and ambient air. Typical settings include:
- Site characterization
- Stationary
source testing and monitoring
- Hazardous waste
sites for determining personal protective equipment (PPE) level
- Fence line
monitoring during removal or remediation activities
- Emergency response
testing
GC analysis with
photoionization detection has been used extensively to characterize and remediate
sites contaminated with volatile organic chemicals (VOCs). Likewise, gas chromatographs
coupled with an electron capture detector are used for analysis on sites contaminated
with chlorinated pesticides. The recent development of truly field portable
quadrupole mass spectrometers now permits GC-mass spectrometry analysis which
provides definitive identification. This instrument and technique is being implemented
during emergency response and counter terrorism situations that requires definitive
identification of contaminants in near real-time.
Because of the
increased durability of modern instruments field gas chromatographs are capable
of the same analyses as fixed laboratory instruments. EPA verified SW846 methods
are then possible with field gas chromatographs and various detectors. Some
SW846 methods include:
- Phenols Method
8041 (GC-FID or ECD
- Phthalates
Method 8061A (GC-ECD)
- Amines Method
8070A (GC-NPD)
- Chlorinated
pesticides Method 8081A (GC-ECD)
- PCBs Method
8082 (GC-ECD)
- PAHs Method
8100 (GC-FID)
- Chlorinated
hydrocarbons Method 8121 (GC-ECD)
- Volatile organic
chemicals (VOC) Method 8240
- Organophosphorus
compounds Method 8141A (GC-NPD or FPD)
- Chlorinated
herbicides Method 8151A (GC-ECD)
- MTBE Method
8260 (GC/MS)
- Halogenated
VOC Method 8260B
- SVOCs/base
neutral acids (BNA) Method 8270C (GC/MS)
- Dioxin Method
8280 (GC/MS)
Theory of Operation
The theory of separation
by GC is relatively simple and understanding the factors that affect separation
allows more effective applications of GC analysis in the field. The purpose
of separation is to allow identification and quantitation of individual components
of a mixture and the theory of separation is detailed below. In addition to
separation, detection of analytes after separation, which is an essential but
separate aspect of chromatography, is presented in the section describing system
components. Basic components of a complete gas chromatographic system include:
(1) a carrier gas supply, (2) a syringe for sample introduction, (3) the injection
port, (4) the column and oven, (5) the detector and data collection system.
Components of a gas chromatograph are presented in greater detail in the section
describing the system components. Schematic diagrams and photographs of instruments
can also be accessed through the Systems Components section.
Before separation
occurs in the chromatographic column, the mixture of components in the sample
is introduced into the chromatograph through the injection
port with a syringe. At this point, the analytes are vaporized (if not already
in the gas phase) by the high temperature maintained in the injection port.
The analytes are kept in the gaseous state by maintaining all elements of the
instrument at a temperature above the boiling point of the analytes. The gas
phase analytes are then immediately swept onto the chromatographic column by
the mobile phase. The mobile phase is comprised of an inert carrier gas, which
usually is nitrogen, helium, or hydrogen.
As
the analytes are swept through the column by the mobile phase, separation occurs based on the affinity of each analyte for the stationary phase. The gas
chromatographic column is composed of a coiled, tubular column and the stationary
phase within the tube. GC columns are either packed or open-tubular. Early GC
columns were packed with carbon or diatomaceous earth based solids which acted
as the stationary phase. In modern open-tubular columns, the stationary phase
is a liquid organic compound that is coated on the internal surface of the fused
silica column. Polarities of the analytes dictate the choice of stationary phase.
Components of the mixture with a high degree of affinity for the stationary
phase are strongly retained while components with low affinity for the stationary
phase migrate rapidly through the column. As a consequence of the differences
in mobility due to affinities for the stationary phase, sample components separate
into discrete bands that can be qualitatively
and quantitatively analyzed.
As individual components
of the mixture elute the chromatographic column, they are swept by the carrier
gas to a detector. The detector generates a measurable electrical signal, referred
to as peaks, that is proportional to the amount of analyte present. Detector
response is plotted as a function of the time required for the analyte to elute
from the column after injection. The resulting plot is called a chromatogram.
Detector response is generally a gaussian shaped curve representative of the
concentration distribution of the analyte band as it elutes from the column.
The position of the peaks on the time axis may serve to identify the components
and the area under the peaks provide a quantitative measure of the amount of
each component.
System
Components
The primary components of a GC include:
(1) injection
port
(2) column
(3) integrator
or data acquisition system
(4) detectors
Other parts include:
(1) autosampler(s)
(2) control
panel, electronic pressure control (EPC)
(3) injection
port liners
(4) septa
(5) ferrules
(6) flow controllers
The carrier gas
is introduced in the injection port where the sample is volatilized and swept
through the column, and where the compounds are separated. The carrier gas/sample
mixture then enters the detector where the compounds are identified. The signal
from the detector then is amplified and displayed by the data system.
A capillary column
is an open tube made of fused silica with an outer coating of durable plastic
and an inner coating of stationary-phase material. Some capillary columns have
a second outer covering of stainless steel to withstand the higher pressure
required to analyze alcohols, ketone, and VOCs by the purge-and-trap method.
A lesser used column type is the packed column. Packed columns use a stainless
steel or glass tube with a 1/8th inch inner diameter packed with a solid stationary
phase.
The effectiveness
of a chromatographic column in separating solutes (analytes) is dependent on
a number of variables. Understanding these variables is essential to the process
of optimizing any chromatographic system and achieving resolution of analytes.
Variables that affect separation include distribution equilibrium constants,
retention time, retention (capacity) factors, and selectivity factors. A discussion
of each of these is presented below.
Distribution Constants
The effectiveness
of a chromatographic column in separating two solutes is dependent on the relative
rates at which the species are eluted. These rates are determined by the degree
the solutes distribute themselves between the mobile and stationary phases.
Distribution equilibria in chromatography defines the transfer of an analyte
between the mobile and stationary phases. The equilibrium constant, referred
to in chromatography as the distribution constant or partition ratio, is the
ratio of the molar concentration of the solute in the stationary phase to its
molar concentration in the mobile phase. This is mathematically expressed as K = cS/cM. If K is constant
over a wide range of solute concentrations, then Cs (concentration
of solute in the stationary phase) is directly proportional to cm (concentration of solute in the mobile phase). When this holds true, chromatographic
peaks are symmetrical, Gaussian distributions and retention times are independent
of the amount of analyte injected.
Retention Times
Retention time
of an analyte is defined as the time it takes after sample injection for the
analyte to elute and reach the detector. The time for unretained species to
reach the detector is defined as the dead time. Click here to see a figure that illustrates each. This “rate of migration” of an unretained
species is the same as the rate of motion of the mobile phase molecule. The
linear rate (v) of a solute molecule is defined as the column length
divided by its retention time and is mathematically expressed v = L/tR where L is the column length and tR is the retention
time of the solute. The linear rate (u) of a mobile phase molecule
is calculated by dividing the column length by the dead time (tM).
Obviously, the
distribution of a solute between the stationary and mobile phases will directly
affect the linear migration rate of a solute in the mobile phase and ultimately
its retention time. The migration rate (v) is expressed as v = u x fraction of time the solute spends in the mobile phase. Thus the
less time a solute spends in the mobile phase, the smaller the fraction to multiply u, and ultimately lowering the migration rate. The lower the migration
rate v, the higher (or longer) the retention time, tr. The fraction
can also be expressed as moles of solute in the mobile phase divided by the
total moles of solute. Total moles of solute in the mobile phase is equivalent
to the concentration of solute times the volume of mobile phase. Likewise, the
total moles of solute is the sum of moles in the mobile and stationary phase.
This is expressed as v = u x (cMVM)
/{(cMVM)+(cSVS)}
or u x 1/{1 +(cSVS)/cSV S}.
Substituting the
equilibrium constant for the ratio of solute concentrations results in v = u x 1/{1+(KVS/VM)},
an equation expressing a solute migration rate as a function of its distribution
constant.
Retention Factor
Another parameter
used to describe migration rates of solutes on gas chromatographic columns is
the retention factor, also referred to as the capacity factor. Mathematically,
the retention factor is expressed as k'=KVS/VM.
By substituting this expression for capacity into the equation for the migration
rate, a relationship between column capacity and migration rates is established.
The relationship is mathematically expressed as v = u x {1/(1+k')}.
When the migration rate, for a solute and mobile phase molecule u are
substituted with values measured from a chromatogram, an equation to determine
the retention factor can be derived. The equation is k'=(tR-tM)/TM.
Selectivity Factors
The ability of
a column to retain one analyte more strongly than a second is a function of
the column's selectivity. The column's selectivity factor for two species, A
and B is mathematically expressed as alpha = KB/KA.
By definition the more strongly retained species is B and therefore A is always
greater than one. Substitution for the distribution constants yields an equation
that allows the selectivity factor alpha to be experimentally determined from
a chromatogram. The equation is alpha = (tR)B - TM/(tR)A TM.
Column Performance
Theoretical
Plates and Height Equivalent Theoretical Plates
While movements
of solutes through a gas chromatographic column are described by distribution
constants, retention times, retention (capacity) factors, and selectivity factors,
column efficiency (performance) is described by a quantitative measure labeled
theoretical plates. The number of theoretical plates (N) is calculated
by dividing the column length (L) by the height equivalent theoretical
plate (H). Plate height is experimentally calculated by dividing the
variance of a Gaussian shaped chromatographic peak divided by the column length.
This is graphically illustrated in this figure. A
separate equation that provides the number of theoretical plates (N)
is N = 16(tR/W)2 and is graphically
illustrated in this figure.
Resolution
A chromatographic
column's ability to separate a mixture of compounds is defined as its resolution.
The mathematical equation for resolution, Rs, is RS = Z/(WA+WB).
Where:
Z = The difference in time between two chromatographic peaks
WA = Peak width (in time units) of Compound A
WB = Peak width (in time units) of Compound B Components
of the equation's units are graphically represented in this figure. As the figure illustrates, the greater the separation of compounds,
the higher the column resolution, RS.
Resolution can
also be calculated using retention factor k' for two solutes, the selectivity
factor, and the number of theoretical plates. The following equation is used
to calculate resolution, or with simple rearrangement, to calculate the number
of theoretical plates required to achieve a desired resolution.
Resolution:
RS = N1/2/4{( -1)/}{k'B/1+k'B}
Theoretical
Plates: N = 16Rs2{ /( -1)}2{(1+k'B)/k'}2
The previous discussions
all assume Gaussian distributions of analytes as they elute from the column.
However, non-gaussian shaped peaks do occur and the peak shapes provide information
relative to chromatographic variables. Two frequent phenomena related to non-gaussian
peak shapes that occur are fronting and tailing. With fronting, the front side
of a peak is drawn out while the tail (or backside) on the right is steep. The
most frequent cause of fronting is too large of a sample introduced into the
column. In contract, the more frequent phenomenon of tailing results in the
right side of a peak (the tail) is drawn out. This usually occurs when the solute
has c concentration dependent (non-linear) distribution coefficient. This can
also be a cause of fronting. The result of fronting and tailing is poor separation
(and thus resolution) and less accurate quantitative analysis.
Optimizing Separations
Stationary
phase - An organic liquid compound that is either coated on or covalently
bonded to the silica surface of a capillary column. Stationary phases are occasionally
solids packed inside the column. The most widely used columns are the fused
silica capillary columns due to strength and flexibility. The polarities of
the compounds of interest dictate the choice of stationary phase, under the
rule “like dissolves like.” Commonly used stationary phases include:
polydimethyl siloxane
(commonly referred to as OV-1 of SE-30) for PCB or PAH, separation carbowax
- used for free acids, alcohols, and glycols
OV-17 - for
pesticides and glycols
OV-210 -
for chlorinated aromatics, nitroaromatics, and alkyl substituted benzenes
OV-3 or SE-52 for
halogenated organics.
Carrier gas - The mobile phase is composed of an inert carrier gas, usually nitrogen, helium,
or hydrogen. The choice of carrier gas is frequently determined by the type
of detector used and subsequent purity requirements. The sample constituents
are transformed into the gaseous phase and are carried along the column during
separation. By increasing the speed (flow rate) of the carrier gas, the analysis
time can be reduced; however, optimal resolution may be compromised. A faster
flow rate also sweeps the injector more efficiently, improving introduction
of the sample into the column. If resolution is not compromised, increased flow
rates can also reduce analysis times.
Length - Capillary columns vary in length from 15 to 100 meters is a coiled configuration
to fit in the instrument oven. For environmental analysis, 30- to 60-meter columns
typically are used. Shortening the length of the column can shorten the analysis
time; however, resolution (separation) will be compromised. Again, if resolution
is not compromised, analysis time can be reduced with a shorter column.
Diameter - Diameters of open tubular capillary columns are typically between 0.32 and
0.25 millimeter, with high resolution columns having diameters of 0.20 to 0.15
millimeter. The smaller diameter columns require special injection splitting
to reduce the sample size and prevent column overload. Columns referred to as
mega bore open tube columns are also available and have a greater capacity but
at the expense of resolution. However, these columns have better resolution
than packed columns. Packed columns have diameters large as 2 millimeters. The
smaller diameter produces better resolution and greater selectivity, but can
handle only a small volume of sample (1 to 2 microliters).
Field analysis
is frequently performed in less than ideal conditions and samples can have concentrations
or constituents that can ruin column performance. When this occurs, columns
are replaced, often at significant expense. At this point, analysts need to
insure that data produced with the new column is consistent with and comparable
to data produced with the previous column. This is accomplished by analyzing
quality control samples to demonstrate comparable separation and sensitivity
is achieved. Numerous other techniques are available to the analyst, such as
the use of internal standards. A discussion of these techniques is beyond the
scope of this text but available in most analytical chemistry text.
Another problem,
associated with coated capillary columns and not exclusive to field analysis,
is column bleed. This phenomenon is best described as the elution of the stationary
phase. A frequent cause of column bleed is excessive oven temperature. The end
result of column bleed is a fouled detector. More sensitive detectors such ac
electron capture detectors are highly susceptible to column bleed.
Gas
Chromatography Detectors
A variety of detectors
for gas chromatographs are available. In general, each detector takes advantage
of a unique characteristic of a molecule and uses that characteristic to generate
a measurable electrical signal. A discussion of the most frequently used detectors
is presented below. Principles of operation, classes of compounds providing
optimal response, and detection limits are included. Click on the detector name
for a schematic of the detector.
Because of the
unique and complex nature of mass spectrometry, the mass spectrometer as a GC
detector is discussed under the mass spectroscopy entry in the encyclopedia.
Click here to go directly
to the mass spectroscopy section.
Note
that the MS detector is the most versatile. The MS is used widely in place of
conventional GC detectors. Standard Operating Procedures (SOPs) are available
for the following GC/MS methods:
Field analysis
is frequently performed in less than ideal conditions and samples can have concentrations
or constituents that can ruin column performance. When this occurs, columns
are replaced, often at significant expense. At this point, analysts need to
insure that data produced with the new column is consistent with and comparable
to data produced with the previous column. This is accomplished by analyzing
quality control samples to demonstrate comparable separation and sensitivity
is achieved. Numerous other techniques are available to the analyst, such as
the use of internal standards. A discussion of these techniques is beyond the
scope of this text but available in most analytical chemistry text.
Another problem,
associated with coated capillary columns and not exclusive to field analysis,
is column bleed. This phenomenon is best described as the elution of the stationary
phase. A frequent cause of column bleed is excessive oven temperature. The end
result of column bleed is a fouled detector. More sensitive detectors such ac
electron capture detectors are highly susceptible to column bleed.
Photoionization
Detector
A photoionization
detector (PID) consists of a special ultraviolet lamp, ranging in energy from
9.5 to 11.7 eV, mounted on a low-volume flow-through cell. As constituents of
the sample pass through the cell, they are energized and ionized. The ions are
collected at positively charged electrodes, where the change in current is measured.
The 10.2 eV lamp
emits ultraviolet light at 121 nanometers (nm), which is sufficient to ionize
BTEX compounds and hexane. A few halogenated compounds that have ionization
potentials of less than 11.7 eV can be detected by the higher-energy PID. The
PID is more selective than the FID.
The PID can detect
VOCs (aromatic and chlorinated) and petroleum constituents including BTEX. The
PID can detect BTEX in the low ppb to high part per trillion range.
The PID is a nondestructive
detector that can be used in series before other detectors. Using multiple detectors
extends the range of compounds that can be detected in one analysis. PID is
sensitive to water and must be recalibrated more often than the FID.
Flame
Ionization Detector
A flame ionization
detector (FID) consists of a stainless steel jet constructed so that carrier
gas exiting the column flows through the jet, mixes with hydrogen, and burns
at the tip of the jet. Hydrocarbons and other molecules which ionize in the
flame are attracted to a metal collector electrode located just to the side
of the flame. The resulting electron current is amplified by a special electrometer
amplifier which converts very small currents to millivolts.
The FID is sensitive
to almost all molecules that contain hydrocarbons. Examples include aromatic
and chlorinated VOCs, petroleum constituents, SVOCs, and PCBs.
The FID is a destructive
detector that can be used in series only after nondestructive detectors. The
FID is sensitive to water, but has a wider linear range of detection than the
PID. The FID also can detect more compounds than the PID.
The
FID can detect compounds that contain the low ppb to high part per trillion
range.
SOPs are available
for the following GC/FID methods:
Electron
Capture Detector
An electron capture
detector (ECD) consists of a sealed stainless steel cylinder that contains radioactive
nickel-63. The nickel-63 emits beta particles (electrons) which collide with
the carrier gas molecules ionizing them in the process. A stable cloud of free
electrons thus forms in the ECD cell. When an electronegative molecule such
as a halogenated molecule enters the cell, it immediately combines with one
of the free electrons which temporarily reduces the number of free electrons.
The detector electronics pulse at a variable rate to measure the electrons remaining
in the cell.
SOPs are available for the following GC/ECD methods:
The ECD is highly
sensitive to electronegative molecules (those capable of producing negatively
charged ions) such as halogenated compounds or those that contain nitrogen.
The ECD readily detects chlorinated pesticides, halogenated solvents, PCBs,
and dioxins.
The ECD is a nondestructive
detector that can be used in series before other detectors. The ECD is sensitive
to water that affects the condition of the Ni-63 foil that covers the detector.
The foil must be reconditioned when its sensitivity diminishes. Because the
ECD contains a radioactive source, users may be subject to licensing requirements.
The Nuclear Regulatory Commission (NRC) is the agency in the United States that
regulates radioactive materials. However, in certain states, the regulations
are enforced by a state agency. There are two types of ECDs that require licenses;
specific license ECDs and general license ECDs. A general license ECD is an
ECD that customers in the United States can purchase without having their own
radioactive material license. General License ECDs are typically covered under
the distributor's distribution license. General License customers are required
to complete and sign the General License Registration Card and comply with regulations.
The owner of a General License ECD becomes a General Licensee when the ECD is
purchased. The owner does not have to apply for a General License from the NRC
or State Agency but could be required to register the ECD within the state.
Specific License ECDs require the end user to have a Materials License from
the NRC or local State Agency, which permits the owner to posses the applicable
type and quantities of radioactive material. Holders of Specific Licenses have
greater flexibility with their ECD and more responsibilities. For example, Specific
License customers are allowed to clean ECDs and work with open sources. Distributors
of ECDs are required to notify customers of licensing requirements.
The ECD can detect
halogenated compounds in the low ppb to part per trillion range. The more halogenated
the molecule, the more sensitive the detector is to that compound. For example,
the ECD is orders of magnitude more sensitive to carbon tetrachloride than to
vinyl chloride.
Electrolytic
Conductivity Detector
An electrolytic
conductivity detector (ELCD) is a halogen-specific detector that operates on electrolytical
conductivity principles. Organic compounds eluding from a GC column form
combustion products as they are mixed with hydrogen gas over a nickel catalyst
at 1,000C in a quartz tube furnace. For example, organic chlorides form hydrochloric
acid (HCl). The HCl readily ionizes and changes the electrolytic conductivity
which is monitored by the ELCD.
Click here for a GC/ELCD method for halogenated volatile organics.
The ELCD is a halogen-specific
detector. The ELCD readily detects chlorinated pesticides, halogenated solvents,
PCBs, and dioxins.
The ELCD is a destructive
detector that can be used in series only after nondestructive detectors. The
ELCD requires a higher degree of maintenance than other detectors. It also has
a wider linear range than the ECD.
The ELCD can detect
halogenated compounds in the low ppb to part per trillion range. The degree
of halogenation has less effect on sensitivity than is the case with the ECD.
Detection limits for compounds are similar, even though the degree of halogenation
may vary.
Thermal
Conductivity Detector
A thermal conductivity
detector (TCD) consists of tiny coiled wires arranged in a wheat stone bridge
configuration. Electric current flows through the filaments making them glow
hot, while carrier gas exiting the column flows past the other two filaments.
The gas flow carries away excess heat, and the filaments equilibrate. When a
sample compound exits the column, the thermal conductivity of the gas flowing
around the filaments is changed. Therefore, the filaments get hotter and the
balance of the wheat stone bridge is altered.
The TCD is used to detect gaseous compounds, such as nitrogen, oxygen, and
other nonhydrocarbon compounds (for example, landfill gases).
The TCD is a destructive
detector that can be used in series only after nondestructive detectors. The
TCD has limited target analyte list. Because the TCD detects nitrogen, nitrogen
cannot be used as a carrier gas.
The TCD can detect
gaseous compounds in the ppm range. TCD is not used for trace analysis. Larger
sample volumes are required to achieve increased sensitivity. To increase sample
volume sufficiently, a large-diameter packed column must be used.
Nitrogen-Phosphorous
Detector
The nitrogen-phosphorus
detector (NPD) is similar to the FID, except that the hydrogen gas flow rate
is reduced to 1 to 3 milliliters per minute (ml/min), and an electrically heated
thermoionic bead is positioned just above the jet orifice. Analyte molecules
exiting the column collide with the hot bead, and the nitrogen or phosphorus
react and liberate an electron. The electron is attracted to the same collector
electrode and electrometer amplifier used in the FID.
The NPD is sensitive
to nitrogen and compounds that contain phosphorus. The NPD typically is used
in analysis for organophosphorus pesticides. In addition, it is used in analysis
of nitroaromatics (explosives).
The NPD is a destructive
detector that can be used in series only after nondestructive detectors. The
NPD is sensitive to water that affects the condition of the thermoionic bead.
The active element on the bead eventually will become depleted and require replacement.
The NPD can detect
nitrogen and compounds that contain phosphorus in the ppb range. Reduced sensitivity
often indicates the depletion of the active element on the thermoionic bead.
Surface
Acoustic Wave Sensors (SAWS)
Surface acoustic
wave sensors (SAWS) are chemical sensors that use a piezoelectric crystal as
a transducer to detect and quantitate individual chemical compounds. A transducer
is defined as an electric device that transmits energy from one system to another.
This transmission of energy frequently occurs by converting the energy to a
different form. Several types of transducers include:
- electrochemical
devices that measures changes in voltage or current
- acoustic devices
that measure changes in frequency as a result of a change in mass bound to
the surface device (SAWS)
- optical devices
that measure changes in fluorescence, light absorbance, or reflectance.
A commercially
available example of a GC/SAW is the Electronic Sensor Technology (EST0 4199),
a handheld, portable (35-pound) gas chromatograph system. The EST 4100 was evaluated
by the U.S. Environmental Protection Agency (EPA) and researchers from Oak Ridge
National Laboratory during an Environmental Technology Verification (ETV) study
in 1997.
The SAWS operates
on the principle of the piezoelectric effect. The piezoelectric effect is defined
as the oscillation of a crystal, at a constant frequency characteristic of the
mass and shape of the quartz crystal, when an appropriate electrical potential
is applied across the face. In general, a SAWS operates by applying an alternating
electric current through a piezoelectric crystal transducer, causing the crystal
to vibrate. As a sample from the GC column effluent is exposed to the sensor
(detector), chemical compounds from the sample are deposited on the crystal
surface, changing the crystal's mass and oscillation frequency.
The piezoelectric
crystal transducer is the critical element of the SAWS detector. The crystal
transducer is a quartz crystal that is coated with a compound that selectively
adsorbs specific molecules. As the GC effluent passes over the transducer, the
coating adsorbs specific compounds and the mass of the crystal transducer increases,
while decreasing the resonant frequency of the crystal. Ultimately, the change
in frequency is correlated to analyte mass. When the analyte desorbs, the crystal
returns to its original frequency.
A SAWS with a chemically
selective coating is designed to respond to a specific compound or compounds,
providing qualitative identification. Other SAWS are non-selective and may respond
to a number of compounds; a GC is used to separate and qualitatively identify
individual compounds. The observed change in acoustic energy from the crystal
provides the quantitative measurement of the concentration of the analyte.
Although the SAWS
is much less sensitive to humidity than to organic compounds, water vapor also
can cause a sensor to respond. Water uptake always will occur because no coating
is perfectly hydrophobic. The use of adsorbants to trap water in a sample before
it is introduced to the sensor can reduce the effects of humidity. Besides chemical
interferences, temperature can have a direct effect on the frequency of oscillation
by causing thermal expansion of the crystal and affecting the acoustic properties
of the chemically selective coating. A technique that often is applied to compensate
for temperature is the use of a second sensor, almost identical to the first,
which is maintained at the same temperature but not exposed to the target chemical.
Target analytes
include:
- volatile organic
compounds (VOCs)
- semivolatile
organic compounds (SVOCs)
- polychlorinated
biphenyls (PCBs).
Minimum detection
limits are approximately 1 to 10 parts per million (PPM) for typical VOCs and
can be reduced to 10 to 100 parts per billion (ppb) through the use of a preconcentrator.
Mode of Operation
The majority of
GC analyses for environmental applications can be divided into two basic groups:
volatile organic chemicals (VOC) and semi-volatile organic chemicals (SVOC).
The division is based essentially on the extraction technique used to separate
the analyte from the sample matrix. VOC analysis relies on relatively low boiling
points and high volatility of the organic chemical. Therefore, the extraction
technique frequently is simply heating the sample matrix and driving the analyte
into the headspace of the sample container. SVOC analysis, however, requires
more sophisticated extraction techniques. SVOC analysis requires solvents to
extract the analytes from the sample matrix. Each of the fundamental techniques
are discussed below, with advantages and limitations of each.
VOC Analysis
Methods of analysis
for VOCs include direct injection for ambient air and soil gas, and static headspace
extraction or purge and trap extraction for soil and water. These techniques
will be discussed in the following slides.
Analytes of interest
include: (1) halogenated VOCs, including vinyl chloride, methylene chloride,
trichloroethene (TCE), tetrachloroethene (PCE), trichloroethane (TCA), chloroform,
carbon tetrachloride, and ethylene dibromide; (2) nonhalogenated VOCs (solvents),
including methyl iso-butyl ketone (MIBK), methyl ethyl ketone (MEK), and acetone;
(3) aromatic compounds, including BTEX and chlorobenzenes; and (4) fuels, including
gasoline, diesel fuel, jet fuel, and kerosene.
Click here for purge and trap sample preparation methods coupled with GC/PID and GC/MS analysis for volatile organics.
Ambient Air:
On site air analysis
can be used to test for VOCs in stationary source testing (emission inventory),
hazardous waste site testing to determine appropriate levels of personal protective
equipment (PPE), fence line monitoring during remediation activities, and emergency
response testing.
The preferred mode
of sample collection for quick analysis is to directly draw a sample of ambient
air into the on site GC or GC/MS using an internal pump in the analytical instrumentation.
Air samples also may be collected in Tedlar bags, Summa® canisters,
on Tenax tubes, or using solid phase microextraction (SPME) devices. Click here for Summa® canister
sample collection methods coupled with GC/FOZ Trap analysis for volatile organics.
The use of these other sampling containers will require that an air sample (1
to 5 ml) be withdrawn from the sample container or desorbed from the sorbent
and injected into the GC system. (Only applies to Tedlar bags and Summa® canisters. Tenax and SPME will need some desorption technique.)
On site analysis
of air generally is conducted using a portable GC system or GC/MS configuration.
Sops for GC methods
for vapor phase organics using field portable instruments include:
Transportable GCs
(larger, lab-grade instruments) can also be used but provide far more logistical
problems. The detection limits for air analysis will range considerably depending
on the method of sample collection and GC system. Typical detection limits will
range from 5 to 200 parts per billion by volume (ppbv). Detection limits will
be considerably lower if the analytes are concentrated on some type of adsorbent
material and desorbed versus directly drawn into the analytical system via a
sampling pump. Analytical times for VOCs should be less than 10 minutes per
sample. A GC/MS system can either be used to provide qualitative to semiquantitative
data in a survey mode or quantitative data in an analytical or selective ion
monitoring (SIM) mode.
The advantages
of on-site air monitoring is the quick data which allows flexibility for on-site
personnel and the project manager.
When conducting
emissions testing, samples that have high moisture or acid content must be pretreated
prior to analysis. If a mass spectrometer is not used, exact analyte identification
(coelution problems) may not be possible in complex mixtures.
Soil Gas:
Soil gas analysis
commonly is used to identify “hot spots” or source areas of VOCs in the subsurface.
It also can be used to approximate the extent of a subsurface plume.
The detection limit
for most VOCs is 10 nanograms per liter (ng/L). Calibrations consist of direct
injection using an air standard mix or methanol-based standards. The concentration
is reported in ng/L. One liter of air weighs approximately 1 gram. Therefore
on a weight basis, ng/L is approximately equivalent to ppb.
Typical sample
containers include glass bulbs, Tedlar bags, Summa® canisters,
syringes or 22 or 40-ml vials. Teflon®-coated syringes, plungers,
or stop-cocks should be avoided, because some VOCs (for example, 1,1,1-TCA)
are strongly sorbed to Teflon® surfaces.
Because analyte
concentrations are generally more dilute in soil gas samples, analysis usually
requires larger sample volumes than liquid samples. Therefore, a 1 to 5 milliliters
aliquot of air from the soil gas sample is injected into the GC column. In contrast,
liquid sample injection volumes are usually a fraction of that volume (microliters)
and contain no air. Because of the relatively large amount of air injected from
soil gas samples, deterioration of the column stationary phase and oxidation
of the ECD foil by oxygen in the sample is possible.
The advantages
of soil gas analysis are that it is rapid, inexpensive, provides real-time results,
and causes minimal disturbance to the site.
One limitation
of soil gas analysis is that it does not always reflect a true soil concentration.
The technique is limited to high volatility and low solubility compounds. Coelution
problems can occur in complex mixtures if a mass spectrometer is not used. Sample
carryover or cross contamination may be a problem in highly concentrated samples.
Decontamination of syringes is critical, especially for chlorinated VOCs.
Static
Headspace:
Static headspace
extraction is widely used in determining VOCs in waste water, soil, and drinking
water. This extraction method is highly productive and cost effective, requiring
minimal sample preparation.
Efficiency of headspace extraction is based on soil or water partition coefficients
of the volatile organic analytes. The principle follows Henry's Law, where the
vapor pressure of the solute in the headspace is proportional to its mole fraction
in solution. In other words, when a sample containing VOCs is sealed in a headspace
vial, the vapor pressure of the VOC in the headspace is proportional to its
concentration in solution. This phenomenon allows for analysis of the headspace
gas to determine VOC concentrations of the sample matrix without time consuming
solvent extractions. Henry's Law is mathematically expressed by the equation
P=HX, where
P = Pressure of
gas above the liquid (atm)
H = Henry's
Law constant (atm-m3/mole)
X = Equilibrium concentration of dissolved gas (mole/m3)
The process can
be simply described by two actions: (1) diffusion of analyte into the headspace
and, (2) diffusion back into the matrix.
A steady state
equilibrium is reached when the concentration in the headspace is equal to the
concentration in the matrix.
Low viscosity liquids
reach equilibrium faster. Solids take longer.
A constant heat
time is recommended for samples that do not reach equilibrium within a reasonable
time.
Analysis time ranges
from 10 to 30 minutes depending on number and boiling point of analytes of interest.
Detection limits
for BTEX compounds and most chlorinated VOCs are in the range of 1 to 10 ppb;
detection limits for TPH-purgeable compounds are 1 to 10 PPM in water and 10
to 50 PPM in soil.
Water samples can
be collected in 40 ml volatile organic analysis (VOA) vials or directly in a
22-ml headspace vial. Soil samples can be collected in a 4-ounce glass jar or
directly into a 22-ml headspace vial. Mass or volume is measured into a headspace
vial (generally 5 ml or 5 grams). No sample preparation is required for water
samples. Sample preparation for soil samples may vary depending on the initial
analyte concentration and soil type. For high concentration soil samples (PPM
levels), studies have shown that methanol-flood achieves the greatest extraction
efficiency. An aliquot of the methanol can then be analyzed using the static
headspace technique. Methanol-flood techniques, however, are not appropriate
for soil samples that contain less than 200 mg/kg of an analyte due to the dilution
effect. For these low concentration soil samples, Method 5021 recommends the
addition of an aqueous matrix modifying solution. For field analysis, the addition
of a matrix modifier may not be necessary due to the quick analysis times.
Listed
below are the steps in the operational sequence of a typical automated
headspace sampler.
The sample is introduced
to the platen heated zone and allowed to equilibrate at a fixed temperature
for a fixed time period. Typical heating temperatures range from 40 to 80C.
Heating times vary from 10 to 30 minutes. After the sample is heated, it is
then mixed to help volatilization into the headspace.
Next, the vial
is raised onto a needle and pressurization gas (nitrogen) fills the vial to
a pressure of 3 to 27 pounds per square inch (psi).
The pressure in
the vial is then allowed to equilibrate for 0.5 to 2.5 minutes.
The vent valve
is then opened and the pressure in the vial displaces the headspace through
the sample loop, filling the loop at the proper loop fill time. Pressure in
the vial will equal atmospheric pressure. Loop volumes can range from 0.1 to
5 ml Typical loop volumes used are 1 or 2 ml
After the loop
fill, the vent valve and pressure valve are closed, allowing the sample vapor
to equilibrate and pressure and flows to stabilize.
When the GC-ready
signal is received by the headspace unit, the 6-port valve rotates and the sample
loop contents are transferred to a heated line with column carrier gas. Carrier
gas then back flushes the loop, sweeping through the heated transfer line into
the GC injection port (0.3 to 6 minutes).
Helium is bubbled
through the solution at ambient temperature and the volatiles are transferred
from the matrix to the vapor phase. The volatiles are then swept through the
sorbent column where they are trapped. Then, the sorbent column is heated and
backflushed with helium gas to desorb the components. The components are then
transferred to a GC via a heated line, where they are separated using the appropriate
column and detected using a mass spectrometer or other detector. Typically,
a 5 ml sample is used for water analysis and a 5 gram sample is used for soil
and sediment analysis.
Purge
and Trap:
The primary advantage
of purge and trap over static headspace is that it is a dynamic process. It
is a more efficient extraction technique for those VOCs which have a higher
octanol/water partition coefficient, especially in soils with high organic matter
content.
Purge and trap
is the recommended VOC extraction technique used for GC analysis. The recommended
extraction methods are 5030A and 5035.
Comparison
of Attributes of Headspace Versus Purge-and-Trap Analysis
The headspace technique
is a static process, while purge-and-trap is a dynamic process. Some of the
differences between the two processes are listed in the table above.
Use of organic
solvents is comparable for the two techniques; however, for purge-and- trap
analysis, water is used as a solvent for soil purging, increasing the amount
of liquid waste produced in the field laboratory.
For headspace analysis,
samples can be collected directly into preweighed headspace vials; no transferring
of sample is required. For purge-and-trap analysis, a measured amount of sample
must be weighed and then transferred into a specially designed purge vessel.
The purge vessels are reusable glassware and must be decontaminated. Use of
such vessels increases the potential for carryover contamination.
Because the headspace
process is static, the headspace can become saturated before equilibrium is
reached between the medium (soil or water) and the headspace concentration.
Therefore, high concentrations of compounds may not be transferred efficiently
to the headspace and subsequently detected. Results for high-concentration samples
may be biased low. In purge-and-trap analysis, compounds can be purged continuously
from the medium and collected onto the trap, allowing higher concentrations
to be transferred and detected effectively.
Because the headspace
system does not allow for concentration of compounds onto a trap over a period
of time (like the purge-and-trap system), trace amounts of compounds that are
not mobilized effectively at equilibrium may not be detected.
Both systems have
the potential for loss of compounds. During pressurization of the headspace
vial, compounds can be lost if the cap is not properly sealed. In the purge-and-trap
system, loss of compounds will occur as a result of volatilization during the
weighing of the sample and its transfer to the purge vessel.
The cost of the
purge-and-trap system is typically more than that of a comparable headspace
system.
The headspace system
is an attachment to the GC that requires little additional bench space; however,
the purge-and-trap system is a separate stand-alone system that may require
twice as much bench space as the GC alone.
The throughput
of the headspace system is significantly larger than that of the purge-and-trap
system, primarily because the purge-and-trap system requires more handling,
transfer, purging, and decontamination of samples than the headspace system.
|
|
|
System process |
Static |
Dynamic |
Solvent use |
Minimal use of solvent |
Minimal use of solvent |
Sample preparation |
Weigh sample in headspace vial |
Weigh sample; transfer to separate purge
vessel |
Sample decontamination |
No transfer vessels to decontaminate |
Reusable purge vessels must be decontaminated |
Upper dynamic range |
Limited because of saturation of headspace
before equilibrium is reached |
Not limited by headspace saturation because
the compounds are purged in a dynamic system |
Detection limit |
No mechanism for concentration of compounds |
Ability to purge for long periods of time
and concentrate compounds onto trap before analysis |
Loss of contaminants |
If headspace vial is not properly crimped,
compounds may be lost during pressurization |
Less loss of compounds because purge vial
is not pressurized;outlet goes directly to trap; however, loss of compounds
may occur during transfer of a sample |
Cost |
Less than that of purge-and- trap system |
More than headspace system |
Space requirements |
Less space than purge-and- trap system |
More space than headspace system and requires
more plumbing |
Throughput |
50 samples per day |
20 samples per day |
SVOC
Analysis
In the past, typical
“formal lab” extraction methods included soxhlet, liquid-liquid,
and sonication.
More recently, the accelerated solvent extractor (ASE) has become the extraction
method of choice because it is more rapid and uses less solvent. Microwave Assisted
Extraction (MAE) is another extraction method that is promising for fixed laboratory
analysis because it too is more rapid and uses less solvent compared to the
“older” solvent extraction methods.
Sops are available
for the following:
Sops for ASE methods
coupled with GC/MS analysis for methyl parathion include:
The key in the
field is to simplify the solvent extraction methods to minimize solvent waste,
save time, and reduce cost. Typical solvents used in the field include hexane,
methanol, methylene chloride, and methyl tert-butyl ether (MTBE). Simplified
field, solvent extraction methods normally do not include a cleanup step. A
concentration step may also be eliminated if elevated detection limits are acceptable.
ASE and MSE are not as commonly used in the field (especially on small projects)
because of the initial expense of the equipment and other logistical constraints
such as power and space requirements.
Extraction techniques
that use a minimal amount of solvent are preferred for field applications. The
techniques listed above fit into this category. Although supercritical
fluid extraction (SFE) is an EPA-approved technique for fixed-laboratory
applications, it has not gained wide use in the field because of the expense
of the apparatus and limited portability. Solid-phase extraction (SPE) and solid-phase
microextraction (SPME) are ideal techniques for field use because they are rapid,
use little or no solvent, simple, and inexpensive. SPE is primarily limited
to water samples, although it can be used as a cleanup technique for liquid
extracts of solid samples. SPME has gained much more popularity in the last
2 to 3 years. Its advantage over SPE is that it can be used for both VOCs and
SVOCs and no solvent is required. Thermal desorption for SVOCs is analogous
to a more rigorous static headspace extraction for VOCs. It also is convenient
for field use because it is simple, rapid, and requires no solvent.
Target Analytes
- Phenols
- Phthalates
- Amines
- Chlorinated
pesticides
- PCBs
- PAHs
- Chlorinated
hydrocarbons
- Volatile organic
chemicals (VOC)
- Organophosphorus
compounds
- Chlorinated
herbicides
- MTBE
- Halogenated VOC
- SVOCs/base
neutral acids (BNA)
- Dioxin
Performance Specs
Performance specs
include information on detection limits, calibration, and quality control.
Detection Limits
Detection limits
for GC analysis are highly dependent on the type of detector used. Part per
million (PPM) and part per billion (ppb) levels are routinely achieved with
most detectors. Detection limits for each respective detector are discussed
in the systems components section.
Part per billion
detection limits have been reported for quadrupole mass spectrometers. Magnetic
sector mass spectrometers, also known as high resolution mass spectrometers,
are significantly more sensitive than the quadrupole. For example, magnetic
sector spectrometers detect dioxins at parts per trillion (ppt) levels in soil
and parts per quadrillion (ppq) in water.
Calibration
Because GC analysis
provides both qualitative and quantitative information, GC calibration requires
that each be addressed.
Qualitative:
Calibration consists
of injecting a known volume of a standard (at a known concentration as well)
and measuring the time between injection and elution. This is known as the retention
time. If all variables such as temperature, flow rate, and column length are
all constant, the retention time of a given compound should remain the same.
To eliminate variables in injection technique, retention time relative to an
internal standard can be used. Thus, if a peak appears in the chromatogram of
a sample at the sample retention time of a standard, the compound is tentatively
identified, not definitively. A chromatogram provides only a single piece of
qualitative information about each species in a sample, its retention time.
It is important to note that while chromatograms may not lead to positive identification
of species present in a sample, they can provide sure evidence of the absence
of a given compound (or is present at a concentration below the detection limit
of the method). A compound can be definitively identified by dual column analysis
of mass spectrometry detection.
Quantitative:
Quantitative gas
chromatography is based on comparison of either peak height or area of the analyte
peak with that of one of the standards. Most modern chromatographs have digital
electronic integrators that calculate accurate peak areas.
The most common
method for quantitative chromatographic analysis involves the preparation of
a series of standard solutions that approximate the composition of the unknown.
Chromatograms for the standards are generated and peak areas are plotted as
a function of concentration. A plot of the data should yield a straight line
passing through the origin. Frequent restandardization is necessary for the
highest accuracy.
The highest precision
for quantitative chromatography is generated by using internal standards to
eliminate uncertainties introduced by sample injection. In this procedure, a
measured quantity of an internal standard is introduced into each standard and
sample. Quantitation is performed by comparing analyte peakness to internal
standard peak areas. It is essential that the internal standard peak be well
separated from the peaks of all other components of the sample.
Quality Control
As stated above,
field-based GC data of the same quality as fixed laboratory data can be generated.
However, to do so, quality control (QC) measures comparable to the fixed laboratory
must be used. Typical QC measures are defined below.
External standard
calibration generally consists of a three-point or five-point calibration.
A medium level
standard (continuing calibration) is analyzed once a day to check the response
of the instrument compared to the average response of the initial calibration.
Method blanks are
analyzed to check for laboratory-induced contamination and instrument blanks
are analyzed to check for contamination induced by the instrument (usually by
sample carryover). Method blanks are especially critical in soil gas analysis
when chlorinated VOCs are target analytes.
Performance evaluation
(PE) samples are spiked matrix samples that have certified concentrations of
analytes and that can be purchased from reputable vendors (for example, Environmental
Resource Associates or Absolute Standards). They are usually analyzed “blind”
by the analyst (meaning the analyst does not know what analytes are present
or their concentrations). The analyst must report the proper analytes and concentrations
within the certified concentration ranges for the laboratory's accuracy to be
acceptable. PE samples are normally not analyzed for soil gas or ambient air
analysis.
Matrix spike/matrix
spike duplicate (MS/MSD) samples are analyzed to check for extraction efficiency
of the analytical system. Percent recoveries are calculated and must fall within
an acceptable range for the extraction efficiency to be acceptable. Percent
recoveries are also compared to each other by calculating relative percent differences
(RPD) to assess the precision of a method. MS/MSDs are not typically run for
soil gas or ambient air analysis.
Laboratory duplicates
are analyzed to assess the precision of the method and homogeneity of the sample.
The laboratory duplicates consist of the analysis of two aliquots of the same
sample. The results from the laboratory duplicates are compared to each other
through RPD calculations.
Surrogate spikes
are necessary to evaluate the extraction efficiency on a per sample basis. A
surrogate compound is one that is chemically similar to the target compounds
but does not coelute with any of them. A percent recovery of the known spiked
concentration is calculated for each sample and compared to site specific control
limits. (Not typically analyzed in soil gas or ambient air samples.)
Laboratory control
samples (LCS) consist of a sample matrix spiked with a known concentration of
standard purchased from a separate vendor other than the one from which the
calibration standards were purchased. The percent recovery for all analytes
must be within an acceptable range for the accuracy of the calibration to be
acceptable.
Dual column analysis
allows confirmation of the compound identity and concentration of a compound.
This is especially important when MS is not used for confirmation.
Advantages
- Coupled with
a mass spectrometer definitive compound identification is possible.
- A broad spectrum
of organic chemical analysis is possible.
- Field portable
gas chromatographic analysis can be of equal quality as fixed laboratory data.
- Dual column
analysis can also provide definitive compound identification.
Limitations
- Coelution of
analytes.
- Instrumentation
operation requires a higher degree of expertise than most other instrumentation.
- Sample carryover
and equipment contamination.
- Multicomponent
compound interferences with individual analytes
- Petroleum
with BTEX or PAH
- PCBs with
chlorinated pesticides
- Toxaphene with chlorinated pesticides
Cost Data
Manufacturers of
gas chromatographs occasionally establish elevated rental prices to encourage
purchase rather than lease. However, scientific instrumentation suppliers such
as HAZCO provide rental gas chromatographs capable of most analyses. Some manufacturers
such as Bruker Daltonics, Inc., also offer a lease to purchase plan.
Purchase price
varies widely depending on instrument capabilities, auto-sampling accessories,
and detectors. Simple chromatographs capable of analyzing simple mixtures of
BTEX can be purchased for less than $10,000. Bench scale instruments with a
mass spectrometer detector can cost over $100,000. Field GC/MS systems can be
purchased for $60,000 to $85,000.
Lease prices also
vary according to capability. Monthly rental for a simple instrument designed
to analyze simple BTEX samples is about $1,500. More sophisticated instruments
lease for approximately $3,000 per month.
Manufacturers listed
below should be contacted directly for cost information.
Verification/Evaluation Reports
Verification of
the performance of site characterization and field analytical technologies is
conducted through a variety of programs. Evaluation and verification reports
from EPA's Superfund Innovative Technologies Evaluation (SITE) Measuring and
Monitoring Program, EPA's Environmental Technology Verification Program (ETV)
program, along with links to certification statements from California EPA's
(CalEPA) California Environmental Technology Certification Program, are provided
below.
Superfund
Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program
The SITE Demonstration Program encourages the development and implementation
of innovative treatment technologies for (1) remediation of hazardous waste
sites and (2) monitoring and measurement. In the SITE Demonstration Program,
the technology is field-tested on hazardous waste materials. Engineering and
cost data on the innovative technologies are gathered so that potential users
can assess the technology's applicability to a particular site. Data collected
during the field demonstration are used to assess the performance of the technology,
the potential need for pre- and post-treatment processing of the waste, applicable
types of wastes and waste matrices, potential operating problems, and approximate
capital and operating costs. The following reports from the measuring and monitoring
program are available for gas chromatography:
No reports available for this technology
EPA's
Environmental Technology Verification (ETV) Program
EPA's Environmental
Technology Verification (ETV) Program verifies the performance of innovative
technologies. ETV was created to substantially accelerate the entrance of new
environmental technologies into the domestic and international marketplaces.
ETV verifies commercialized, private sector technologies. After the technology
has been tested, the companies will receive a verification report that they
can use in marketing their products. The results of the testing also are available
on the Internet. The following reports from the ETV program are available for gas chromatography:
- Bruker-Franzen
Analytical Systems, Inc. Model EM640TM was verified for measurement
of volatile organics in soil, water, and soil gas. The verification documents
available consist of a verification report and verification statement.
- Electronic
Sensor Technology Model 4100 Model was verified for measurement of chlorinated
volatile organic compounds in water. The verification documents available
consist of a verification report and verification statement.
- Inficon, Inc.
- HAPSITE with Headspace Sampling Accessory was verified for measurement of
chlorinated volatile organic compounds in water. The verification documents
available consist of a verification report and verification statement.
- Photovac (Perkin-Elmer
Corporation - Photovac Monitoring Instruments) Voyager was verified for measurement
of chlorinated volatile organic compounds in water. The verification documents
available consist of a verification report and verification statement.
- Sentex Systems
Scentograph Plus II was verified for measurement of chlorinated volatile organic
compounds in water. The verification documents available consist of a verification
report and verification statement.
- Viking SpectraTraksTM 672 was verified for measurement of volatile organics in soil, water, and
soil gas. The verification documents available consist of a verification
report and verification statement.
California
EPA's California Environmental Technology Certification Program
CalEPA's environmental technology certification program
is a voluntary program that provides participating technology developers,
manufacturers, and vendors an independent, recognized third-party evaluation
of the performance of new and mature environmental technologies. Developers
and manufacturers define quantitative performance claims for their technologies
and provide supporting documentation; CalEPA reviews that information and,
when necessary, conducts additional testing to verify the claims. The technologies,
equipment, and products that are proven to work as claimed are given official
state certification. The certification program is voluntary and self-supporting.
Companies participating in the program pay the costs of the evaluation and
certification of their technologies. Technologies that have been certified
through this program are listed below. Links are provided to the web sites
that provide the Certified Environmental Technology Transfer Advisory and
Certification Notice for the technologies.
No reports available for this technology
Disclaimer
http://clu-in.org/char/technologies/gc.cfm
Page Last Modified: November 30, 2006

|